PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4811580 · Received June 2, 2015

Report

Report Number
2032227-2015-17676
Event Type
Malfunction
Date Received
June 2, 2015
Date of Event
May 15, 2015
Report Date
May 15, 2015
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A BUTTON ERROR AND HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE WAS FOUND INSIDE OF THE INSULIN PUMP. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON THE DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL THAT THEIR BUTTONS OF THE INSULIN PUMP WERE NOT RESPONDING. THE BLOOD GLUCOSE READING IS 349 MG/DL. THE INSULIN PUMP WILL BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
355469 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAH

Patients

Seq Age Sex Outcome Treatment
1 34 YR