FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4811315 · Received June 2, 2015

Report

Report Number
3004209178-2015-61349
Event Type
Malfunction
Date Received
June 2, 2015
Date of Event
May 11, 2015
Report Date
May 11, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. THE INSULIN PUMP RECEIVED WITH BROKEN BATTERY TUBE THREAD, CRACKED RESERVOIR TUBE LIP, CRACKED CASE NEAR DISPLAY WINDOW CORNER, AND STAINED END CAP STICKER NOTED. NO MOISTURE DAMAGE ON ELECTRONICS ASSEMBLY NOTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY RECEIVED A BUTTON ERROR ALARM THAT THEY WERE UNABLE TO CLEAR. THE CUSTOMER'S BLOOD GLUCOSE WAS 11 MMOL/L. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER STATED THEY MAY HAVE EXPOSED THE INSULIN PUMP TO MOISTURE FROM SWEAT. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED. CUSTOMER AGREED TO RETURN THE PRODUCT FOR ANALYSIS.