FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4811241 · Received June 2, 2015

Report

Report Number
3004209178-2015-61467
Event Type
Malfunction
Date Received
June 2, 2015
Date of Event
May 12, 2015
Report Date
May 12, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED DISPLACEMENT TEST AND MOTOR ERROR ALARM DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. THE INSULIN PUMP WAS UNABLE TO VERIFY ALARM IN THE ALARM HISTORY DUE TO MOTOR ERROR ALARM DURING REWIND. THE INSULIN PUMP RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON LCD WINDOW AND SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED VIA PHONE CALL, THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 285 MG/DL. TROUBLESHOOTING OCCURRED. ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP, AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
356406 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 61 YR