530G INSULIN PUMP
Report
- Report Number
- 3004209178-2015-61309
- Event Type
- Malfunction
- Date Received
- June 2, 2015
- Date of Event
- May 11, 2015
- Report Date
- May 11, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO NUMBERS RAMPING UP NOTED. THE INSULIN PUMP HAS CRACKED CASE AT THE DISPLAY WINDOW CORNER AND CRACKED RESERVOIR TUBE LIP.
IT WAS REPORTED VIA PHONE CALL, THAT THE NUMBERS ON THE INSULIN PUMP SCROLL, WITHOUT ANY INPUT FROM THE CUSTOMER. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 368 MG/DL. TROUBLESHOOTING OCCURRED. ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP, AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 356821 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |