PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4811200 · Received June 2, 2015

Report

Report Number
3004209178-2015-61295
Event Type
Injury
Date Received
June 2, 2015
Date of Event
May 11, 2015
Report Date
May 11, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NV
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE PUMP HAD MINOR SCRATCHES ON THE LCD WINDOW AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED MOTOR ERROR TWICE DURING A BOLUS ATTEMPTS. THE CUSTOMER STATED THAT HE DID THE REWIND AND THREE HOURS LATER HE RECEIVED ANOTHER MOTOR ERROR ALARM. THE DEVICE WAS NOT EXPOSED TO A MAGNETIC FIELD. CUSTOMER DOES NOT USE THE SENSOR FEATURE AND IS ABLE TO COMPLETE THE REWIND SEQUENCE. BLOOD GLUCOSE VALUE WAS 470 MG/DL. CUSTOMER WAS ADVISED TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO A BACK A PLAN. CUSTOMER WAS ADVISED THE INSULIN PUMP WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
356682 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 32 YR