FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4811081 · Received June 2, 2015

Report

Report Number
3004209178-2015-61232
Event Type
Malfunction
Date Received
June 2, 2015
Date of Event
May 11, 2015
Report Date
May 11, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

NO REWIND ANOMALY NOTED. THE INSULIN PUMP PASSED REWIND TEST, BASIC OCCLUSION TEST AND DISPLACEMENT TEST. THE INSULIN PUMP WAS RECEIVED WITH MOTOR ERROR ALARM DURING OCCLUSION TEST DUE TO FAULTY FORCE SENSOR. THE MOTOR WAS TESTED OUTSIDE OF THE INSULIN PUMP AND PASSED. THE INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND SHIFTED END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THE INSULIN PUMP ALARMED MOTOR ERROR DURING BOLUS/BASAL. THE DEVICE WAS NOT EXPOSED TO HIGH MAGNETIC FIELD. ALARM HAS OCCURRED ONCE IN THE PAST MONTH. CUSTOMER DOESN'T USE THE SENSOR FEATURE AND WAS UNABLE TO REWIND THE DEVICE. CUSTOMER WAS ADVISED THE DEVICE NEED TO BE REPLACED AND AGREED TO RETURN THE DEVICE FOR FURTHER ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
356518 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554WWS

Patients

Seq Age Sex Outcome Treatment
1