AUTOCAT2 WAVE
Report
- Report Number
- 1219856-2015-00114
- Event Type
- Malfunction
- Date Received
- June 1, 2015
- Date of Event
- May 12, 2015
- Report Date
- May 12, 2015
- Manufacturer
- ARROW INTERNATIONAL INC.
- Product Code
- DSP
- PMA / PMN Number
- K060309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4).
(B)(4). EVALUATION: THE POWER SUPPLY (P/N: 77-0063-001, S/N: (B)(4)) WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF POWER SUPPLY WAS PERFORMED AND NO ABNORMALITY WAS NOTED. THE PCS ASSEMBLY (P/N: 96-3006-001, S/N: (B)(4)) WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF PCS ASSEMBLY WAS PERFORMED AND FOUND NO OBVIOUS DAMAGE OR DEFECTS. THE POWER SUPPLY WAS INSTALLED INTO A KNOWN GOOD AUTOCAT2W AND FUNCTIONAL TESTING WAS PERFORMED. THE PUMP WAS STARTUP AS NORMAL. A POWER SUPPLY VOLTAGE CHECK WAS PERFORMED AND ALL VOLTAGES WERE WITHIN SPECIFICATION. HOWEVER, A WHISTLE SOUND FOR NOTED AT POWER DOWN. THE PUMP WAS RUN ON BATTERY (BATTERY LOAD TEST) UNTIL THE IABP SHUT DOWN (>90MINUTES WITHIN SPEC). A BATTERY LOAD TEST WAS PERFORMED AFTER LETTING THE PUMP CHARGE FOR OVER 8 HOURS AND THE UNIT RAN FOR OVER 90 MINUTES. THE POWER SUPPLY PASSED FUNCTIONAL TESTING. VISUAL INSPECTION OF POWER SUPPLY INTERNAL HARDWARE WAS PERFORMED AND NO ABNORMALITY WAS FOUND. SEE OTHER REMARKS SECTION. OTHER REMARKS: THE PCS ASSEMBLY IN QUESTION WAS INSTALLED INTO KNOWN GOOD AUTOCAT2W AND PERFORMED FUNCTIONAL TESTING. THE KNOWN GOOD AUTOCAT2W WITH THE PCS ASSEMBLY IN QUESTION INSTALLED PASSED FUNCTIONAL TEST. NO ALARMS WERE OCCURRED DURING THE FUNCTIONAL TEST. THE PCS ASSEMBLY IN QUESTION WAS THEN INSTALLED ONTO THE LEAK TESTER AND PASSED LEAK TESTING. A POPPING NOISE WAS NOTED TO BE COMING FROM RELIEF VALVE OF THE PCS ASSEMBLY DURING PURGE AND DRAIN CYCLES. HOWEVER, THIS FINDING DID NOT AFFECT A FUNCTIONALITY OF THE PCS ASSEMBLY. THE PCS ASSEMBLY PASSED FUNCTIONAL TESTING. VISUAL INSPECTION OF PCS ASSEMBLY INTERNAL HARDWARE WAS PERFORMED AND NO ABNORMALITY WAS FOUND. A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE LOT NUMBER/SERIAL NUMBER WITH NO RELEVANT FINDINGS. THE DEVICE PASSED ALL MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. CONCLUSION: THE REPORTED COMPLAINTS OF "EXCESSIVE NOISE AND WHISTLING NOISE ON POWER DOWN" ARE CONFIRMED. THE REPORTED PROBLEMS WERE REPLICATED DURING FUNCTIONAL TESTING. THE WHISTLING NOISE WAS NOTED FROM THE POWER SUPPLY ON POWER DOWN, AND ALSO EXCESSIVE NOISE WAS FOUND TO BE COMING FROM THE RELIEF VALVE OF THE PCS ASSEMBLY. HOWEVER, THESE FINDINGS DO NOT AFFECT FUNCTIONALITY OF THE POWER SUPPLY AND PCS ASSEMBLY. THE POWER SUPPLY AND PCS ASSEMBLY BOTH PASSED FUNCTIONAL TESTING.
IT WAS REPORTED VIA A FIELD SERVICE REPORT: (B)(4) SYMPTOM. THE (IABP) INTRA-AORTIC BALLOON PUMP WAS MAKING EXCESSIVE NOISE DURING USE ON PATIENT, PUMP WAS SWITCHED WITH ANOTHER IAP-0500. THE BIOMED ALSO REPORTED LOUD WHISTLING NOISE ON POWER DOWN. FINDINGS/ACTION TAKEN: FOUND PUMP MAKING WHISTLING NOISE ON POWER DOWN. FOUND PUMP MAKES POPPING NOISE DURING PURGE AND DRAIN CYCLES. REPLACED WAS THE POWER SUPPLY FOR WHISTLING NOISE ON POWER DOWN. FOUND POPPING NOISE COMING FROM PCS ASSEMBLY, REPLACED PCS ASSEMBLY. UNIT HAD PASSED FUNCTIONAL TESTING PRIOR TO REPAIR. UNIT PASSED (WWS) WORLD WIDE SUPPORT FUNCTIONAL CHECKOUT AFTER REPAIR; REPLACED POWER CORD THAT CUT IN THE INSULATION. FCN LEVEL: 14, SOFTWARE LEVEL: 2.23.
IT WAS REPORTED VIA A FIELD SERVICE REPORT: (B)(4). SYMPTOM: THE (IABP) INTRA- AORTIC BALLOON PUMP WAS MAKING EXCESSIVE NOISE DURING USE ON PATIENT, PUMP WAS SWITCHED WITH ANOTHER IAP-0500. THE BIOMED ALSO REPORTED LOUD WHISTLING NOISE ON POWER DOWN. FINDINGS / ACTION TAKEN: FOUND PUMP MAKING WHISTLING NOISE ON POWER DOWN. FOUND PUMP MAKES POPPING NOISE DURING PURGE AND DRAIN CYCLES. REPLACED WAS THE POWER SUPPLY FOR WHISTLING NOISE ON POWER DOWN. FOUND POPPING NOISE COMING FROM PCS ASSEMBLY, REPLACED PCS ASSEMBLY. UNIT HAD PASSED FUNCTIONAL TESTING PRIOR TO REPAIR. UNIT PASSED (WWS) WORLD WIDE SUPPORT FUNCTIONAL CHECKOUT AFTER REPAIR. REPLACED POWER CORD THAT CUT IN THE INSULATION. FCN LEVEL: 14. SOFTWARE LEVEL: 2.23.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 351775 | AUTOCAT2 WAVE | IABP AUTOCAT 2 WAVE PUMP CONSOLE | DSP | ARROW INTERNATIONAL INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INTRA-AORTIC BALLOON CATHETER |