FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 4810571 · Received May 29, 2015

Report

Report Number
1314492-2015-06409
Event Type
Malfunction
Date Received
May 29, 2015
Date of Event
May 1, 2015
Report Date
May 6, 2015
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPEC IN RELATION TO THE SYSTEM ERROR 105, WHICH WAS REPRODUCED DURING EVAL. SYSTEM ERROR 105 WAS CONFIRMED AND CAUSED BY A SEIZED MOTOR. THE FAILED MOTOR ASSEMBLY WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 105. IT WAS ALSO REPORTED THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350824 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1