FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 4810532 · Received May 29, 2015

Report

Report Number
1314492-2015-06396
Event Type
Malfunction
Date Received
May 29, 2015
Date of Event
May 1, 2015
Report Date
May 6, 2015
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

MFR REF NO: (B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPEC IN RELATION TO THE SYSTEM ERROR 105, WHICH WAS NOT REPRODUCED. SYSTEM ERROR 105 WAS CONFIRMED THROUGH A REVIEW OF THE EVENT HISTORY LOG AND CAUSED BY A SEIZED MOTOR. THE FAILED MOTOR ASSEMBLY WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 105. IT WAS ALSO REPORTED THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349539 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700