FDA Adverse Event Malfunction Summary report: N

INTROCAN SAFETY

MDR report key: 4810520 · Received May 28, 2015

Report

Report Number
9610825-2015-00202
Event Type
Malfunction
Date Received
May 28, 2015
Date of Event
November 8, 2014
Report Date
May 14, 2015
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
FOZ
PMA / PMN Number
K021094
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL DEVICE WAS NOT RETURNED FOR EVAL. WITHOUT THE ACTUAL SAMPLE, A THOROUGH INVESTIGATION COULD NOT BE PERFORMED. THE BATCH MFG RECORD WAS REVIEWED AND THERE WERE NO SUCH DEFECT ENCOUNTERED DURING IN-PROCESS INSPECTION AND AT FINAL CONTROL INSPECTION. THE PROCESS CARDS SHOWED NO ABNORMALITIES. A HISTORICAL REVIEW OF THE COMPLAINT DATABASE IDENTIFIED NO ADVERSE TRENDS FOR PROD CODE 4552543-02 OR THE INVOLVED LOT. PER THE EVENT DESCRIPTION, IT WAS REPORTED THAT "WHEN THE USED SUPPLIES WERE BEING PICKED UP THE NURSE WAS STUCK BY THE USED NEEDLE". IT SHOULD BE NOTED THAT THE INTROCAN SAFETY IS DESIGNED TO REDUCE THE RISK OF NEEDLE STICK INJURIES. HOWEVER, CDC GUIDELINES AND/OR FACILITY PROTOCOLS SHOULD ALWAYS BE FOLLOWED. SHARPS SHOULD BE DISPOSED IF IMMEDIATELY INTO AN APPROPRIATE SHARPS CONTAINER. IF A SAMPLE AND/OR ADD'L PERTINENT INFO BECOME AVAILABLE, A F/U REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

AS REPORTED BY THE USER FACILITY THROUGH MEDWATCH REPORT: "EVENT DESC: RN SUCCESSFULLY STARTED AN IV THEN PUSHED THE NEEDLE COVER TO SECURE DEVICE. WHEN THE USED SUPPLIES WERE BEING PICKED UP THE NURSE WAS STUCK BY THE USED NEEDLE. UPON INSPECTION IT APPEARED THAT THE NEEDLE COVER MALFUNCTIONED AND DIDN'T COVER THE SHARP. NO HARM TO THE PT. NURSE WAS EXPOSED TO PT'S BLOOD/BODY FLUIDS VIA UNINTENTIONAL NEEDLE STICK. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
346292 INTROCAN SAFETY IV SAFETY CATHETER FOZ B. BRAUN MELSUNGEN AG 14F26G8304

Patients

Seq Age Sex Outcome Treatment
1