FDA Adverse Event Injury Summary report: N

MEGADYNE

MDR report key: 4810474 · Received June 1, 2015

Report

Report Number
8020045-2015-00043
Event Type
Injury
Date Received
June 1, 2015
Date of Event
April 28, 2015
Report Date
May 28, 2015
Manufacturer
LEONHARD LANG GMBH
Product Code
GEI
PMA / PMN Number
K063161
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RETAINED SAMPLES OF THE SAME LOT HAVE BEEN INSPECTED VISUALLY AND TESTED ELECTRICALLY. MECHANICAL TESTS WERE PERFORMED ON 2 RETAINED SAMPLES. ALL SAMPLES WERE FOUND TO PERFORM WITHIN LIMITS. NO FAULTS WERE DETECTED. AS THE INJURY WAS ON THE BUTTOCK CHEEKS AND NOT UNDERNEATH THE DISPERSIVE ELECTRODE, WE ASSUME THAT THE BURN WAS CAUSED BY A DIRECT ELECTRICAL CONTACT BETWEEN THE PT AND THE OPERATING ROOM TABLE DUE TO THE PRESENCE OF MOISTURE THERE (RESIDUAL DAMPNESS AFTER URINATING EVENT) AND NOT BY THE DISPERSIVE ELECTRODE. AFTER SEVERAL REQUESTS TO OBTAIN ADDITIONAL INFO, WE GOT INFORMED BY OUR CUSTOMER "THAT THE PERCEIVED SKIN ISSUES HAD COMPLETELY RESOLVED WITHIN 24 HOURS". THEY HAVE CLOSED THE COMPLAINT. NO INFO WAS MADE AVAILABLE TO US ON HOW THE INJURY WAS TREATED. BASED ON THE RECEIVED INFO, WE WILL CLOSE OUT THE INVESTIGATION.

Description of Event or Problem · 1

ON (B)(6) WE HAVE BEEN INFORMED ABOUT "A (B)(6) LB PT WAS GETTING A T AND A PROCEDURE AT (B)(6) MEDICAL CENTER, (B)(6) OUTPATIENT SURGERY DEPT. ON (B)(6) (...) WHICH RESULTED IN TWO PT BURNS. THE PT WAS PLACED ON TOP OF AN OPERATING ROOM TABLE DRAPED WITH LINEN SHEET, BLANKET DRAW SHEET. M2K-01 USED WITH FORCE FX GENERATOR. THE PT URINATED PRIOR TO THE ONSET OF THE PROCEDURE. NURSING DRIED PT AND PLACED FOLDED UP BLANKET, DISPOSABLE DRAPE AND A DISPOSABLE IMPERVIOUS ABSORBENT PAD UNDER PT. PT DID NOT HAVE CATHETER OR DIAPER. UPON INITIATION OF ELECTROSURGERY THERE WAS NOT ENOUGH POWER AT THE PENCIL SO CIRCULATOR DISCONNECTED MS #0845 AND APPLIED 0865C MEGADYNE PAD (WEIGHT LIMIT UP TO 33LBS). SURGERY PROCEEDED. MS WAS UNPLUGGED. AT THE END OF THE PROCEDURE IT WAS NOTED THAT THE PTS BUTTOCKS WAS RED AND ONE SMALL BLISTER WAS PRESENT ON EACH BUTTOCK CHEEK. THE ABSORBENT PAD AND BLANKET UNDER THE PT WERE DAMP. CIRCULATOR NURSE INDICATED THE BLISTERS WERE NOT THERE AT THE SKIN ASSESSMENT PRIOR TO SURGERY." NO FURTHER DETAILS HAVE BEEN SPECIFIED DESPITE REPEATED REQUESTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353276 MEGADYNE ELECTROSURGICAL DISPERSIVE ELECTRODE GEI LEONHARD LANG GMBH RSW22A30 40220-0744

Patients

Seq Age Sex Outcome Treatment
1 UNK Other