FDA Adverse Event Malfunction Summary report: N

CONSTELLATION VISION SYSTEM

MDR report key: 4810348 · Received June 2, 2015

Report

Report Number
2028159-2015-06245
Event Type
Malfunction
Date Received
June 2, 2015
Date of Event
May 8, 2015
Report Date
July 9, 2015
Manufacturer
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K101285
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE SYSTEM WAS EXAMINED AND THE REPORTED EVENT WAS NOT REPLICATED BUT WAS CONFIRMED (VIA EVENT LOGS). A COMPANY REPRESENTATIVE RESEATED THE SUPERVISOR COMMUNICATION CABLES AS A PREVENTATIVE MEASURE. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY.

Additional Manufacturer Narrative · 1

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT RECEIVED TO DATE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT DURING SURGERY A SYSTEM MESSAGE WAS DISPLAYED AND THE PATIENT´S EYE DEFLATED DUE TO THE EVENT. THE PROCEDURE WAS COMPLETED. THE PATIENT IMPACT WAS UNKNOWN AT THE TIME OF THIS REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT RECEIVED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
355616 CONSTELLATION VISION SYSTEM UNIT, PHACOFRAGMENTATION HQC ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER TABLETOP NA

Patients

Seq Age Sex Outcome Treatment
1 Other