MICRO VASCULAR PLUG
Report
- Report Number
- 2029214-2015-00594
- Event Type
- Malfunction
- Date Received
- June 1, 2015
- Date of Event
- April 8, 2014
- Report Date
- April 10, 2014
- Manufacturer
- REVERSE MEDICAL CORP/COVIDIEN
- Product Code
- KRD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
NO MODEL NUMBER OR LOT NUMBER WAS REPORTED FOR THE DEVICE. THE PRODUCT WAS NOT RETURNED BACK TO REVERSE MEDICAL CORP. THE MICRO VASCULAR PLUG REMAINS IMPLANTED WITHIN THE PATIENT. THE USER USED THE DEVICE FOR TEMPORARY OCCLUSION WHICH THE DEVICE IS NOT INDICATED FOR. THE DEVICE WAS USED WITH EMBOLIZATION PARTICLES AND GLUE WHICH IT IS NOT INDICATED FOR. THE PHYSICIAN BELIEVES THAT IT WAS USER ERROR. THE USER USED THE MICRO VASCULAR PLUG IN A MANNER WHICH WAS NOT CONSISTENT WITH THE INDICATIONS FOR USE.
MEDTRONIC COVIDIEN/REVERSE MEDICAL CORP RECEIVED INFORMATION THAT DURING A BRAIN ARTERIOVENOUS MALFORMATION EMBOLIZATION TREATMENT, THE PHYSICIAN REPORTED THAT THE MVP GOT DETACHED UNINTENTIONALLY. THE PHYSICIAN OPENED AN MVP ON THE VENOUS SIDE TO STOP THE FLOW AND TO ENABLE THE PHYSICIAN TO INJECT EMBOLIZATION PARTICLES ON THE ARTERIAL SIDE. THE PROCEDURE WENT AS PLANNED. THE PHYSICIAN ALSO INJECTED SOME GLUE TO COMPLETE THE EMBOLIZATION PROCEDURE. AT THE MOMENT OF RETRIEVING THE MVP, IT GOT DETACHED. IT HAD REMAINED OPEN IN THE VEIN FOR ABOUT 1 HOUR. THE PHYSICIAN SUGGESTED THAT THE DETACHMENT POINT GOT WEAKENED BY BEING LEFT FOR A LONG TIME IN THE BLOOD VESSEL. IT WAS REPORTED THAT INSTEAD OF RESHEATHING THE MVP BY PUSHING THE MICROCATHETER OVER THE PLUG, THE PHYSICIAN PULLED ON THE PLUG. NO PATIENT INJURY WAS REPORTED AS A RESULT OF THIS PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 353502 | MICRO VASCULAR PLUG | DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION | KRD | REVERSE MEDICAL CORP/COVIDIEN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |