FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4808545 · Received May 28, 2015

Report

Report Number
1052693-2015-00803
Event Type
Malfunction
Date Received
May 28, 2015
Date of Event
April 29, 2015
Report Date
December 17, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER HAD AN INACCURATE REFERENCE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. PRODUCT CODES UPDATED (68 CODE NO LONGER AVAILABLE).

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT NOT YET RETURNED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF ABOUT HIGH BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 120-135MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 05/31/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY. CUSTOMER WAS UNABLE TO CONFIRM WHEN THE TEST STRIPS WERE FIRST OPENED. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 186MG/DL AND 177MG/DL FASTING. REVIEWED METER MEMORY: (B)(6). NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 120-135MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 05/31/2017. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY. CUSTOMER WAS UNABLE TO CONFIRM WHEN THE TEST STRIPS WERE FIRST OPENED. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 186MG/DL AND 177MG/DL FASTING. REVIEWED METER MEMORY: 1:206MG/DL (B)(6) 2015 09:42:59 AM FASTING: YES; 2:405MG/DL (B)(6) 2015 09:42:59 AM FASTING: YES; 3:163MG/DL (B)(6) 2015 09:42:59 AM FASTING: YES; 4:182MG/DL (B)(6) 2015 09:42:59 AM FASTING: YES; 5:179MG/DL (B)(6) 2015 09:42:59 AM FASTING: YES. NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344769 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PR1988

Patients

Seq Age Sex Outcome Treatment
1 0 YR