FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4808495 · Received May 28, 2015

Report

Report Number
1052693-2015-00825
Event Type
Malfunction
Date Received
May 28, 2015
Date of Event
May 4, 2015
Report Date
December 18, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION TEST STRIPS. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Additional Manufacturer Narrative · 1

PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).

Description of Event or Problem · 1

CONSUMER COMPLAINT OF RECEIVING AN E3 ERROR MESSAGE. UPON TROUBLESHOOTING HIGH CONTROL TEST RESULTS WERE OBTAINED. CUSTOMER STATES THAT HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER WAS NOT SURE WHAT HIS EXPECTED RESULT SHOULD BE. VERIFIED THE STRIPS EXPIRE 02/28/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY. REVIEWED METER MEMORY: 148 MG/DL, (B)(6) 2015, 09:30 AM, FASTING: YES; 132 MG/DL, (B)(6) 2015, 09:30 AM, FASTING: YES; 124 MG/DL, (B)(6) 2015, 09:30 AM, FASTING: YES; 121 MG/DL, (B)(6) 2015, 09:30 AM, FASTING: YES; 142 MG/DL, (B)(6) 2015, 09:30 AM, FASTING: YES. NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF RECEIVING AN E3 ERROR MESSAGE. UPON TROUBLESHOOTING HIGH CONTROL TEST RESULTS WERE OBTAINED. CUSTOMER STATES THAT HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER WAS NOT SURE WHAT HIS EXPECTED RESULT SHOULD BE. VERIFIED THE STRIPS EXPIRE 02/28/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY. REVIEWED METER MEMORY: 1:148MG/DL (B)(6) 2015 09:30:00 AM FASTING:YES. 2:132MG/DL (B)(6) 2015 09:30:00 AM FASTING:YES. 3:124MG/DL (B)(6) 2015 09:30:00 AM FASTING:YES. 4:121MG/DL (B)(6) 2015 09:30:00 AM FASTING:YES. 5:142MG/DL (B)(6) 2015 09:30:00 AM FASTING:YES. NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344770 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PS2231

Patients

Seq Age Sex Outcome Treatment
1 0 YR