BRASSERLER USA MEDICAL, LLC MILLING HANDPIECE
Report
- Report Number
- 2025102-2015-00002
- Event Type
- Malfunction
- Date Received
- May 29, 2015
- Date of Event
- May 6, 2015
- Report Date
- May 28, 2015
- Manufacturer
- BRASSELER USA, MEDICAL L.L.C.
- Product Code
- HSZ
- PMA / PMN Number
- K935367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
FOLLOW UP: THE HANDPIECE WAS EVALUATED AND WAS IDENTIFIED TO BE IN EXTREMELY POOR CONDITION, SHOWING EXCESSIVE CORROSION AND RUST. THE HANDPIECE WAS SOLD TO CUSTOMER IN (B)(6) OF 2008. THERE IS NO RECORD OF THIS HANDPIECE BEING RETURNED TO BRASSELER FOR SERVICE/PREVENTIVE MAINTENANCE. IT APPEARS THAT THE HANDPIECE HAS NOT RECEIVED ANY PREVENTIVE MAINTENANCE. THE OPERATING INSTRUCTIONS FOR USE RECOMMENDS PREVENTIVE MAINTENANCE EVERY 6 MONTHS. THIS FAILURE IS ATTRIBUTED TO AN EXTENDED LAPSE OF SERVICE/PREVENTIVE MAINTENANCE.
SURGERY WAS BEING DONE AT (B)(6) SURGERY CENTER. PATIENT WAS PREPARED FOR THE PFJ PROCEDURE AND THE INCISIONS WERE MADE. REPRESENTATIVE OPENED DISPOSABLE INSTRUMENTS (SCREWS AND BURR) FOR SURGERY. ATTEMPT WAS MADE TO INSERT BURR INTO HANDPIECE (LOT 08010104) AND IT COULD NOT BE DONE. IT APPEARS THAT A BURR HAD BROKEN OFF IN A PREVIOUS SURGERY. THE BROKEN PIECE CANNOT BE PULLED OUT OF THE HANDPIECE. THE SECOND HAND PIECE (LOT 08010207) WAS BROUGHT OUT TO BE USED. THIS HANDPIECE IS JAMMED AND DOES NOT TURN. PATIENT HAD TO BE CLOSED UP AND MOVED TO (B)(6) HOSPITAL. SURGERY WAS COMPLETED ON (B)(6) 2015.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350039 | BRASSERLER USA MEDICAL, LLC MILLING HANDPIECE | SURGICAL MILLING HANDPIECE | HSZ | BRASSELER USA, MEDICAL L.L.C. | ZIMMER MILL 00-5927-040-00 REV E |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |