BONE PIN- 3.2 X 140MM- STERILE (2 PACK)
Report
- Report Number
- 3005985723-2015-00088
- Event Type
- Injury
- Date Received
- May 29, 2015
- Date of Event
- May 19, 2015
- Report Date
- May 19, 2015
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K112507
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: DURING PARTIAL KNEE REPLACEMENT SURGERY, A 3.2X140MM BONE PIN BROKE OFF AND EMBEDDED IN PATIENT AS SURGEON ATTEMPTED DRILL THE PIN INTO THE PATIENT. THE BROKEN PIECE OF BONE PIN WAS REMOVED FROM THE FEMUR AND THE CASE CONTINUED WITHOUT INCIDENT. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION WAS NOT PERFORMED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. DIMENSION INSPECTION COMPLETED AT THE TIME OF MANUFACTURING SHOWS THE LOT WAS WITHIN SPECIFICATION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES (111621 NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED AND SHIPPED TO MILLSTONE ON 11/11/2014 AND ACCEPTED INTO FINAL STOCK ON 11/11/2014 PER ERP. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN TRACKWISE AND CATSWEB RELATED TO P/N 111620, LOT NUMBER W38525 SHOWS NO COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. TRACKING OF COMPLAINTS RELATED TO THE 111620 PART NUMBER WILL BE TRACKED THROUGH QUARTERLY TREND REQUEST #(B)(4). CONCLUSIONS: AS PART OF THE INVESTIGATION INTO THIS COMPLAINT, THE STRYKER REP THAT WAS PRESENT FOR THE CASE WAS CONTACTED. IT WAS DETERMINED THAT THE SURGEON DID NOT USE THE DRILL GUIDE DURING PLACEMENT OF THE PINS. DUE TO THE NON-USE OF A DRILL GUIDE AS REQUIRED PER MAKOPLASTY PARTIAL KNEE APPLICATION USER GUIDE, 206388 REVISION 01 FOR PLACEMENT OF BONE PINS, THIS EVENT CONSTITUTES AN OFF-LABEL USE OF THE DEVICE. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD. DEVICE WAS NOT RETURNED FOR EVALUATION.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. BONE PIN DISCARDED.
A SURGEON WAS PERFORMING A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE WHILE THE SURGEON WAS DRILLING A BONE PIN INTO THE PATIENT'S FEMUR, THE BONE PIN BROKE. THE BONE PIN WAS REMOVED FROM THE PATIENT AND THE OUTCOME OF THE CASE WAS SUCCESSFUL.
A SURGEON WAS PERFORMING A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE WHILE THE SURGEON WAS DRILLING A BONE PIN INTO THE PATIENT'S FEMUR, THE BONE PIN BROKE. THE BONE PIN WAS REMOVED FROM THE PATIENT AND THE OUTCOME OF THE CASE WAS SUCCESSFUL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350083 | BONE PIN- 3.2 X 140MM- STERILE (2 PACK) | STEREOTAXIC DEVICE, ACCESSORY | OLO | MAKO SURGICAL CORP. | W38525-2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention |