NSK X95
Report
- Report Number
- 9611253-2015-00026
- Event Type
- Injury
- Date Received
- May 26, 2015
- Date of Event
- November 21, 2013
- Report Date
- May 19, 2015
- Manufacturer
- NAKANISHI, INC.
- Product Code
- EGS
- PMA / PMN Number
- K972563
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- SERVICE PERSONNEL
Narratives
UPON RECEIPT OF THE DEVICE INVOLVED IN THE MDR EVENT, NAKANISHI CONDUCTED A FAILURE ANALYSIS OF THE RETURNED DEVICE THAT INCLUDED MEASUREMENT OF THE TEMPERATURE OF THE OPERATING (C131129-02-1). THESE ACTIVITIES ARE DESCRIBED IN MORE DETAIL BELOW: METHODOLOGY USED: NAKANISHI EXAMINED THE DEVICE HISTORY RECORD FOR THE SUBJECT X95 DEVICE (SERIAL NUMBER (B)(4)). THERE WERE NO PROBLEMS OBSERVED DURING THE MANUFACTURING OR TESTING NOTED IN THE DHR. NAKANISHI CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE AND PERFORMED A SIMPLE MOVEMENT TEST. THERE WERE NO VISIBLE ABNORMALITIES, SUCH AS CRACKS OR DENTS, ON THE OUTSIDE OF THE HANDPIECE. THE DEVICE WAS CONFIGURED FOR TEMPERATURE TESTING IN THE EXACT STATE IN WHICH IT WAS RETURNED. SPECIFICALLY, NO LUBRICATION OR CLEANING WAS PERFORMED ON THE RETURNED DEVICE BEFORE THIS FIRST TEST IN ORDER TO CHARACTERIZE THE DEVICE UNDER CONDITIONS THAT WOULD DUPLICATE THE USE SITUATION AT THE TIME OF EVENT. TEMPERATURE SENSORS WERE FIRST ATTACHED TO THE EXTERIOR OF THE DEVICE AT TEST POINT (E.G., MOST PROXIMAL TO THE PATIENT AND ALONG POINTS FARTHER TOWARD THE DISTAL END OF THE DEVICE). THE TEST SETUP WAS PREPARED TO TAKE TEMPERATURE MEASUREMENTS AT ALL POINTS SIMULTANEOUSLY, INCLUDING A REFERENCE MEASUREMENT AT AMBIENT ROOM TEMPERATURE. WITH WATER AND COOLANT AIR: NAKANISHI ATTACHED A THERMOCOUPLE (SENSOR TO MEASURE A TEMPERATURE) TO EACH OF THE TESTING POINTS FOR THE HANDPIECE TO MEASURE THE EXOTHERMIC SITUATION WITH WATER SPRAY AND COOLANT AIR. NAKANISHI MEASURED THE TEMPERATURE RISE OF THE RETURNED HANDPIECE SET AT 200,000 RPM (MOTOR REVOLUTION 40,000 RPM. NAKANISHI CONFIRMED THAT ONE OF THE TEST POINTS REACHED 48 DEGREES C WITHIN 50 SECONDS AFTER BEGINNING OF THE MEASUREMENT, THEN THE TEMPERATURE WENT DOWN TO 40 DEGREES C. WITHOUT WATER AND COOLANT AIR: NAKANISHI ATTACHED A THERMOCOUPLE (SENSOR TO MEASURE A TEMPERATURE) TO EACH OF THE TESTING POINTS FOR THE HANDPIECE TO MEASURE THE EXOTHERMIC SITUATION WITHOUT WATER SPRAY AND COOLANT AIR. NAKANISHI MEASURED THE TEMPERATURE RISE OF THE RETURNED HANDPIECE SET AT 200,000 RPM (MOTOR REVOLUTION 40,000 RPM). NAKANISHI CONFIRMED THE TEMPERATURE AT EACH OF THE TEST POINTS AS FOLLOWS AFTER THE BEGINNING OF THE MEASUREMENT; IT ROSE TO 70 DEGREES C SUDDENLY. NAKANISHI DISCONTINUED THE MEASUREMENT IN ORDER TO PRESERVE THE HANDPIECE. IDENTIFICATION OF THE SPECIFIC FAILURE MODE(S) AND/OR MECHANISM(S) AND THE ASSOCIATED DEVICE COMPONENT(S) INVOLVED: NAKANISHI DISASSEMBLED THE HANDPIECE AND PERFORMED A VISUAL INSPECTION OF THE INSIDE PARTS. NAKANISHI DID NOT OBSERVE ANY BROKEN PARTS, BUT OBSERVED SOME DIRT AND INSUFFICIENT OIL THAT MIGHT HAVE CAUSED THE HEAT-UP OF THE SUBJECT HANDPIECE. CONCLUSIONS REACHED BASED ON THE INVESTIGATION AND ANALYSIS RESULTS: THE SUBJECT HANDPIECE WAS HEATED UP BECAUSE OF THE DIRT ON THE CARTRIDGE INSIDE THE HANDPIECE ALONG WITH THE USAGE WITHOUT WATER SPRAY AND COOLANT AIR. INSUFFICIENT MAINTENANCE OF THE HANDPIECE CAUSED DIRT ON THE CARTRIDGE. ADDITIONALLY, THE USER DID NOT USE THE HANDPIECE ACCORDING TO NAKANISHI USER MANUAL. THE COMBINATION OF THESE FACTORS CAUSED THE HEAT-UP OF THE SUBJECT HANDPIECE.
THIS MDR IS BEING REPORTED AT THIS TIME AS PART OF OUR INTERNAL REVIEW OF PAST COMPLAINTS AND SERVICE RECORDS. DUE TO THE INCIDENT BEING IN THE PAST, WE ARE LIMITED IN THE INFORMATION WE CAN OBTAIN FROM THE INITIAL COMPLAINANT. THE FOLLOWING INFORMATION IS FROM THE DISTRIBUTOR TO NAKANISHI INC. (NSK) REGARDING A DEVICE MANUFACTURED BY NSK. EVENT SUMMARY: ON (B)(4) 2013, NSK RECEIVED AN ELECTRIC DENTAL HANDPIECE MODEL X95, SERIAL NUMBER (B)(4) FOR REPAIR FROM A DISTRIBUTOR REPAIR FACILITY. ON THE DISTRIBUTOR'S DOCUMENTATION WAS STATEMENT: PATIENT BURNED WITH THIS HANDPIECE. NAKANISHI DID NOT GET ANY DETAILED INFORMATION. COMPLAINT REVIEW: THERE WERE NO PREVIOUS REPAIRS ON FILE FOR THIS HANDPIECE. INVESTIGATION: THE EVALUATION FINISHED ON 12/09/2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 339902 | NSK X95 | EGS | NAKANISHI, INC. | X95 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |