FDA Adverse Event Malfunction Summary report: N

COMPACT AIR DEVICE (CAD II)

MDR report key: 4803311 · Received May 28, 2015

Report

Report Number
8030965-2015-10323
Event Type
Malfunction
Date Received
May 28, 2015
Report Date
September 11, 2014
Manufacturer
SYNTHES OBERDORF
Product Code
HWE
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. (B)(4) EVALUATED THE DEVICE AND OBSERVED THAT THE GEARS WERE BROKEN AND TORN OFF . THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WEAR FROM NORMAL USE AND SERVICING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT THE COMPACT AIR DRIVE (CAD II) DEVICE DID NOT FUNCTION. IT WAS FURTHER REPORTED THAT AIR WAS ESCAPING FROM THE DEVICE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
345470 COMPACT AIR DEVICE (CAD II) INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE SYNTHES OBERDORF NA

Patients

Seq Age Sex Outcome Treatment
1