FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 4802947 · Received May 28, 2015

Report

Report Number
1416980-2015-23225
Event Type
Malfunction
Date Received
May 28, 2015
Date of Event
May 14, 2015
Report Date
May 14, 2015
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE CASSETTE WAS NOT RETURNED, A DEVICE ANALYSIS COULD NOT BE COMPLETED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET/LINE) ALARM OCCURRED DURING DWELL FIVE OF SIX OF PERITONEAL DIALYSIS THERAPY ON THE HOMECHOICE. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THERE WAS NOTHING FOUND DURING TROUBLESHOOTING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ALARM. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344930 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE S15A26055

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE