FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 4802524
·
Received May 26, 2015
Report
- Report Number
- 1314492-2015-06297
- Event Type
- Malfunction
- Date Received
- May 26, 2015
- Date of Event
- May 1, 2015
- Report Date
- May 1, 2015
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Removal / Correction Number
- Z-1484-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER REC'D AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPEC IN RELATION TO THE REPORTED SYSTEM ERROR 322 WHICH WAS NOT REPRODUCED. SYSTEM ERROR 322 WAS CONFIRMED THROUGH REVIEW OF THE EVENT HISTORY LOG. THE SOFTWARE WAS UPGRADED TO CORRECT THIS ISSUE PER THE APPROVED REMEDIAL ACTION ACTIVITIES. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE. BAXTER HAS INITIATED A CAPA TO FURTHER INVESTIGATE THIS ISSUE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341215 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |