FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 4802497 · Received May 26, 2015

Report

Report Number
1314492-2015-06259
Event Type
Malfunction
Date Received
May 26, 2015
Date of Event
April 1, 2015
Report Date
April 30, 2015
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN REC'D BY BAXTER FOR EVAL. THE EVAL HAS NOT BEEN COMPLETED AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE PROVIDED WHEN THE INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP DELIVERED MORE THAN THE PROGRAMMED AMOUNT DURING VOLUMETRIC TESTING. USING WATER AS THE FLUID, THE DELIVERY RATE WAS PROGRAMMED AT 200 ML/HR FOR 15 MINUTES. AFTER THE 15 MINUTES, IT WAS STATED THAT >100 ML WAS COLLECTED. IT WAS ALSO REPORTED THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341214 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1