FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 4802471 · Received May 26, 2015

Report

Report Number
1314492-2015-06293
Event Type
Malfunction
Date Received
May 26, 2015
Date of Event
May 1, 2015
Report Date
May 1, 2015
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Removal / Correction Number
Z-1484-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYSTEM ERROR 322 WHICH WAS NOT REPRODUCED. SYSTEM ERROR 322 WAS CONFIRMED THROUGH REVIEW OF THE EVENT HISTORY LOG. EVALUATION FOUND THE UPPER LINK SWITCH SLOW TO RESPOND. THE FAILED UPPER AUXILIARY ASSEMBLY WAS REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE. BAXTER HAS INITIATED A CAPA TO FURTHER INVESTIGATE THIS ISSUE. THE SOFTWARE WAS UPGRADED PER THE APPROVED REMEDIAL ACTION ACTIVITIES.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PATIENT INJURY.