FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4802428
·
Received May 28, 2015
Report
- Report Number
- 2032227-2015-17137
- Event Type
- Malfunction
- Date Received
- May 28, 2015
- Date of Event
- May 9, 2015
- Report Date
- May 9, 2015
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. THE INSULIN PUMP HAS MINOR SCRATCHES ON THE LCD WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNERS, BELT CLIP SLOT AND BROKEN RESERVOIR TUBE LIP.
Description of Event or Problem · 1
CUSTOMER REPORTED VIA PHONE CALL THAT THERE WAS A BUTTON ERROR ALARM. THE BLOOD GLUCOSE READING IS 167 MG/DL. THE INSULIN PUMP WILL BE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 346243 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |