LUMINOS AGILE
Report
- Report Number
- 2240869-2015-12031
- Event Type
- Malfunction
- Date Received
- May 26, 2015
- Date of Event
- May 7, 2015
- Report Date
- May 7, 2015
- Manufacturer
- SIEMENS MEDICAL SOLUTIONS USA, INC.
- Product Code
- JAA
- PMA / PMN Number
- K111292
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE REPORTED FOOTBOARD WAS REQUESTED FOR FURTHER INVESTIGATION. THE REPORTED ISSUE IS UNDER INVESTIGATION AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION HAS BEEN RECEIVED. THIS REPORT WAS SUBMITTED (B)(6) 2015.
THE INVESTIGATION OF THE RETURNED FOOTREST SHOWED BREAKAGE OF THE STEERING RAIL. THIS BREAKAGE HAS NO SAFETY RELATED IMPACT ON THE ATTACHMENT OF THE FOOTREST AND ITS HALT. IT IS ASSUMED THAT THE BREAKAGE IN THE RAIL WAS CAUSED BY AN IMPROPERLY ATTACHED FOOTREST AND FOLLOWING TILTING. THE DETENT AND THE BOLTS OF THE FOOTREST ARE COMPLETELY FUNCTIONAL. MOREOVER ALL SAFETY MARKS ARE VISIBLE. THEREFORE THE INVESTIGATED PART DID NOT SHOW ANY DEFECTS REGARDING FUNCTIONALITY AND RELIABILITY OF THE FOOTREST. IT IS ASSUMED, THAT THE FOOTREST FELL OFF BECAUSE IT WAS NOT PROPERLY ATTACHED BY THE USER. CORRECT ATTACHMENT OF THE FOOTREST IS DESCRIBED IN THE OPERATOR MANUAL XPD1-320.620.01.02.02, CHAPTER 8, PAGES 73 AND 74. THE FOOTREST WAS REPLACED AT THE CONCERNED CUSTOMER SITE. THIS REPORT WAS SUBMITTED APRIL 1, 2016.
IT WAS REPORTED THAT A FOOTBOARD OF THE LUMINOS AGILE SYSTEM BECAME LOOSE DURING THE EXAM. THE TABLE WAS TILTED TO APPROXIMATELY 30 DEGREES WITH A PATIENT STANDING ON THE FOOTBOARD WHEN IT BECAME DISLODGED AND THE PATIENT SLID OFF. NO INJURIES WERE SUSTAINED AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340003 | LUMINOS AGILE | SYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-I | JAA | SIEMENS MEDICAL SOLUTIONS USA, INC. | 10502200 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |