FDA Adverse Event
Injury
Summary report: N
PRECISION SPECTRA®
MDR report key: 4802390
·
Received May 28, 2015
Report
- Report Number
- 3006630150-2015-01353
- Event Type
- Injury
- Date Received
- May 28, 2015
- Date of Event
- May 11, 2015
- Report Date
- May 11, 2015
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- 030017
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE EXPLANTED DEVICE WAS NOT RETURNED TO BSN AS IT WAS KEPT BY THE MEDICAL FACILITY. IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORDS WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MED WATCH WILL BE FILED.
Description of Event or Problem · 1
A REPORT WAS RECEIVED THAT THE PATIENT LOST STIMULATION RIGHT AFTER A FALL. THE LEAD HAD CONTACTS THAT WERE DAMAGED. AN X-RAY WAS TAKEN AND CONFIRMED LEAD FRACTURE. THE LEAD WAS REPLACED. THE PATIENT WAS REPORTEDLY DOING WELL POSTOPERATIVELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 347360 | PRECISION SPECTRA® | SPINAL CORD STIMULATOR | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-8216-70 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Required Intervention |