FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4802361 · Received May 28, 2015

Report

Report Number
3004209178-2015-59994
Event Type
Malfunction
Date Received
May 28, 2015
Date of Event
May 7, 2015
Report Date
May 7, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH A LOOSE AND PROTRUDED DRIVE SUPPORT CAP, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND A MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE DRIVE SUPPORT CAP ON THE INSULIN PUMP STARTED GETTING LOOSE AND THEN CAME OFF. THE CUSTOMER STATED THAT THE INSULIN PUMP MIGHT HAVE BEEN DROPPED OR BUMPED TO THE FLOOR BUT WAS NOT SURE. BLOOD GLUCOSE VALUE WAS 80 MG/DL. CUSTOMER WAS ADVISED TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO A BACK A PLAN. CUSTOMER WAS ADVISED THE INSULIN PUMP WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
345733 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 46 YR