FDA Adverse Event Malfunction Summary report: N

BRAVO

MDR report key: 4801933 · Received May 26, 2015

Report

Report Number
9710107-2015-00135
Event Type
Malfunction
Date Received
May 26, 2015
Report Date
May 21, 2015
Manufacturer
GIVEN IMAGING LTD.
Product Code
FFT
PMA / PMN Number
K002028
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO REQUIRED INTERVENTION TO PREVENT PERMANENT IMPAIRMENT/DAMAGE IN THIS CASE. THE INFORMATION WAS UPDATED IN THIS SUPPLEMENTAL. EVALUATION SUMMARY: ONE DELIVERY SYSTEM AND CAPSULE WERE RETURNED TO MEDTRONIC FOR EVALUATION. THE DELIVERY SYSTEM WAS INVESTIGATED VISUALLY FOR EXTERNAL DAMAGE. THE TROCAR NEEDLE WAS ADVANCED. THE DELIVERY SYSTEM WAS NOT BENT AND THE PLUNGER WAS NOT BROKEN. THE EMERGENCY RELEASE ON THE DELIVERY SYSTEM WAS NOT IMPLEMENTED. THE CAPSULE ELECTRODES WERE VISUALLY ACCEPTABLE. THE WIRE THAT HOLDS THE CAPSULE WAS COMPLETELY OFF AND THE FOAM GASKET WAS IN GOOD CONDITION. THE DELIVERY SYSTEM DID NOT HAVE ANY VISIBLE DAMAGE. AS THE PRODUCT WAS RECEIVED, THE DEVICE FUNCTIONED PER SPECIFICATION.

Description of Event or Problem · 1

CUSTOMER REPORTED ON A BRAVO PH CAPSULE THAT FAILED TO ATTACH. THE PATIENT WAS NOT INJURED FOLLOWING THE PROCEDURE.

Description of Event or Problem · 1

A REPEAT PROCEDURE WAS NECESSARY DUE TO THE ALLEGED DEVICE MALFUNCTION. INTERVENTION WAS NOT REQUIRED. THERE WAS NOTHING UNUSUAL ABOUT THE PATIENT OR PROCEDURE ITSELF THAT MAY HAVE LED TO THIS EVENT. AN ENDOSCOPY WAS PERFORMED PRIOR TO THE PROCEDURE AND THE ESOPHAGUS APPEARED TO BE NORMAL. THE DEVICE OPERATOR HAS BEEN PERFORMING THIS PROCEDURE FOR OVER FIVE YEARS. NO KNOWN ADVERSE EVENTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340451 BRAVO BRAVO PH CAPSULE FFT GIVEN IMAGING LTD. FGS-0313 26700Q

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention