FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4801353 · Received May 28, 2015

Report

Report Number
3004209178-2015-59936
Event Type
Malfunction
Date Received
May 28, 2015
Date of Event
May 7, 2015
Report Date
May 7, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED UNABLE TO PRIME DURING THE PRIME/A33 TEST AND MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED LOOSE DRIVE SUPPORT DISK. THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP HAS MINOR SCRATCHED LCD WINDOW, CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS CRACKED, BROKEN BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP AND RESERVOIR TUBE CRACKED.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

A CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THEIR INSULIN PUMP WAS SQUIRTING OUT OF THEIR INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER STATES THAT THE INSULIN PUMP WAS STUCK IN THE PREPARING TO PRIME LOOP. THE CUSTOMER STATES THAT THE RESERVOIR AND TUBING CONNECTOR ARE DRY. THE CUSTOMER WAS INFORMED THAT THE PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. A REPLACEMENT PUMP WAS SHIPPED TO THE CUSTOMER. THE CUSTOMER SENT THE PRODUCT BACK FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347664 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 48 YR