FDA Adverse Event Malfunction Summary report: N

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 4800654 · Received May 28, 2015

Report

Report Number
9611451-2015-00259
Event Type
Malfunction
Date Received
May 28, 2015
Date of Event
April 29, 2015
Report Date
April 30, 2015
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) THE RT268 INFANT EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K103767 METHOD: THE COMPLAINT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE AND WAS VISUALLY INSPECTED AND PRESSURE TESTED. RESULTS: VISUAL INSPECTION REVEALED A CRACK IN THE YELLOW SLE RESTRICTOR, ALONG THE WELD LINE FROM THE FEMALE END OF THE CONNECTOR. THE PRESSURE TEST REVEALED THAT THE CIRCUIT WAS WITHIN SPECIFICATION AND DID NOT EXHIBIT SIGNIFICANT LEAK. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT NUMBER 141217. WE HAVE ONLY RECEIVED ONE OTHER COMPLAINT OF THIS NATURE FOR THE RT268 BREATHING CIRCUIT. CONCLUSION: THE CRACK ALONG THE FLOW LINE COULD HAVE MADE IT WEAKER THAN USUAL, AND IT THUS CRACKED WHEN THE USER TIGHTENED THE CONNECTORS DURING SET UP. THIS WAS LIKELY CAUSED BY A MOLDING DEFECT IN THE RESTRICTOR. ALL INFANT BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE AND FLOW TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT268 STATE THE FOLLOWING: CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE. PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. SET APPROPRIATE VENTILATOR ALARMS.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THEY FOUND A CRACK IN THE RESTRICTOR OF AN RT268 INFANT EVAQUA 2 BREATHING CIRCUIT. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347169 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT268 141217

Patients

Seq Age Sex Outcome Treatment
1