FDA Adverse Event Malfunction Summary report: N

SPINE REFERENCE CLAMP

MDR report key: 4800351 · Received May 27, 2015

Report

Report Number
1723170-2015-00643
Event Type
Malfunction
Date Received
May 27, 2015
Date of Event
May 1, 2015
Report Date
February 23, 2016
Manufacturer
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
Product Code
HAW
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

CORRECTION: IT WAS DISCOVERED ON 27-JAN-2016, THAT AN INCORRECT DATE WAS REFERENCED IN A PREVIOUS SUPPLEMENTAL 3500A SUBMISSION. THE REPORTED DATE OF 16-AUG-2016 WAS REPORTED INCORRECTLY AND SHOULD BE 16-AUG-2015.STATEMENT IN PREVIOUS SUPPLEMENTAL 3500A SUBMISSION SHOULD READ: ¿ON 30-JUN-2015, IT WAS NOTICED THAT A CODING ERROR IN MDR SUBMISSIONS FROM OUR FACILITY RESULTED IN THE MAUDE DATABASE INCORRECTLY CODING THE DEVICES RELATED TO OUR MDR SUBMISSIONS FROM 25-MAY-2015 TO 16-AUG-2015. THE DECISION TO WAIT UNTIL THE DATABASE WAS CORRECTED WAS MADE AFTER CONSULTATION WITH THE FDA AS ADVISED BY A CONSUMER SAFETY OFFICER WITH THE INFORMATION ANALYSIS BRANCH, DIVISION OF POST MARKET SURVEILLANCE, OFFICE OF SURVEILLANCE AND BIOMETRICS. AN IT SOLUTION WAS IMPLEMENTED ON 16-AUG-2015. THIS MDR WAS SUBMITTED TO CORRECT THE CODING ERROR. THERE IS NO NEW INFORMATION TO CHANGE THE PATIENT INFORMATION, EVENT DESCRIPTION AND/OR MANUFACTURER NARRATIVE THAT WAS PREVIOUSLY REPORTED.¿

Additional Manufacturer Narrative · 1

PATIENT AGE NOW PROVIDED. PATIENT SEX NOW PROVIDED. MEDTRONIC INVESTIGATION OF RETURNED SUSPECT DEVICE FINDS THAT ONE OF THE TEETH ON THE CLAMP HAS BROKEN OFF AND IS MISSING. ANOTHER TOOTH IS BENT. OTHERWISE, THE CLAMP IS IN GOOD WORKING CONDITION. THE REPORTED EVENT WAS CONFIRMED TO BE CAUSED BY THE PHYSICAL DAMAGE OF THE CLAMP TEETH. THE HARDWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS RELATED TO A HARDWARE ISSUE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION HARDWARE ANOMALY TRACKING DATABASE.

Additional Manufacturer Narrative · 1

ON 13-OCT-2015, IT WAS NOTICED THAT A PREVIOUS MDR SUBMISSION CONTAINED INCORRECT INFORMATION WITH REGARDS TO THE COMMON DEVICE NAME, PRODUCT CODE AND/OR PMA/510(K). THIS MDR IS BEING SUBMITTED TO CORRECT THIS INFORMATION. THERE IS NO NEW INFORMATION TO CHANGE THE PATIENT INFORMATION, EVENT DESCRIPTION OR MANUFACTURER NARRATIVE THAT WAS PREVIOUSLY REPORTED.

Additional Manufacturer Narrative · 1

PATIENT INFORMATION WAS NOT MADE AVAILABLE FROM THE SITE. RETURN REQUESTED. NO PARTS HAVE BEEN RECEIVED BY MANUFACTURER FOR ANALYSIS. PART NOT RECEIVED BY MANUFACTURER.

Description of Event or Problem · 1

A MEDTRONIC REPRESENTATIVE REPORTED THAT, WHILE IN A SPINE PROCEDURE, A SITE REPORTED THAT THEIR CERVICAL CLAMP WAS DAMAGED, ONE CLAW WAS BENT. IT WAS NOTED THAT THE SURGEON USED IT WITH THIS KNOWLEDGE, AND NAVIGATION WENT FINE. NO FURTHER DETAILS REGARDING THE DAMAGE, OR HOW IT OCCURRED, WERE PROVIDED. THE SURGEON COMPLETED THE PROCEDURE WITH THE USE OF THE NAVIGATION SYSTEM. THERE WAS NO IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343064 SPINE REFERENCE CLAMP NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. (LOUISVILLE) NONE 140307

Patients

Seq Age Sex Outcome Treatment
1 80 YR