VENTED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2015-00249
- Event Type
- Malfunction
- Date Received
- May 27, 2015
- Date of Event
- April 24, 2015
- Report Date
- April 28, 2015
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED A VERTICAL CRACK IN THE CHAMBER DOME BESIDE THE HINGE BRACKET STRETCHING TO THE BASE. RESIDUE WAS PRESENT ON THE CHAMBER. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT 140604. CONCLUSION: THE RESIDUE SUGGEST THAT THE DAMAGE WAS CAUSED BY THE CHAMBER COMING INTO CONTACT WITH A SOLUTION CONTAINING ETHANOL/ALCOHOL WHICH HAS RESULTED IN ENVIRONMENTAL STRESS CRACKING OF THE CHAMBER DOME. EVERY MR290 CHAMBER IS PRESSURE TESTED TO 200 CMH2O FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER STATE THE FOLLOWING: - DO NOT SOAK, WASH, STERILIZE, OR REUSE THIS PRODUCT. AVOID CONTACT WITH CHEMICALS, CLEANING AGENTS, OR HAND SANITIZERS. - SET APPROPRIATE VENTILATOR ALARMS. - PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. - MAXIMUM OPERATING PRESSURE: 8 KPA. (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT THE CHAMBER DOME OF AN MR290V HUMIDIFICATION CHAMBER WAS FOUND CRACKED AFTER FOUR DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341866 | VENTED HUMIDIFICATION CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 140604 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |