FDA Adverse Event Malfunction Summary report: N

TRUETRACK

MDR report key: 4795161 · Received May 21, 2015

Report

Report Number
1052693-2015-00773
Event Type
Malfunction
Date Received
May 21, 2015
Date of Event
April 25, 2015
Report Date
January 4, 2016
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K032657
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION.

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE. PRODUCT CODES UPDATED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS VIA HUSBAND. EXPECTED BLOOD GLUCOSE RESULTS FASTING RANGE FROM 100 TO 120 MG/DL. COMPARING BLOOD GLUCOSE RESULTS FROM HIS TRUETRACK METER (B)(4) TO WIFE'S TRUETRACK METER (B)(4) AND OBSERVED 60 TO 80 MG/DL DIFFERENCE. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. BACK TO BACK BLOOD TEST DECLINED. VERIFIED STORAGE OF TEST STRIPS IS WITHIN INSTRUCTED SPECIFICATION. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 02/28/2017 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: 312 MG/DL, (B)(6) 2015, 09:09 AM; 161 MG/DL, (B)(6) 2015, 08:00 AM; 131 MG/DL, (B)(6) 2015, 09:30 AM; 130 MG/DL, (B)(6) 2015, 07:00 AM; 128 MG/DL, (B)(6) 2015, 09:04 AM. ADVERSE EVENT NOT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS VIA HUSBAND. EXPECTED BLOOD GLUCOSE RESULTS FASTING RANGE FROM 100 TO 120MG/DL. COMPARING BLOOD GLUCOSE RESULTS FROM HIS TRUETRACK METER (B)(4) TO WIFE'S TRUETRACK METER (B)(4) AND OBSERVED 60 TO 80MG/DL DIFFERENCE. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. BACK TO BACK BLOOD TEST DECLINED. VERIFIED STORAGE OF TEST STRIPS IS WITHIN INSTRUCTED SPECIFICATION. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 02/28/2017 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED FASTING FROM METER MEMORY: 312MG/DL (B)(6) 2015 09:09AM; 161MG/DL (B)(6) 2015 08:00AM; 131MG/DL (B)(6) 2015 09:30AM; 130MG/DL (B)(6) 2015 07:00AM; 128MG/DL (B)(6) 2015 09:04AM; ADVERSE EVENT NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
333782 TRUETRACK BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUETRACK RR4516

Patients

Seq Age Sex Outcome Treatment
1 0 YR