LIFEVEST WCD 3100 SYSTEM
Report
- Report Number
- 3008642652-2015-03306
- Event Type
- Injury
- Date Received
- May 20, 2015
- Date of Event
- April 11, 2015
- Report Date
- May 15, 2015
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THERE WAS NO DEATH OR DEVICE MALFUNCTION ASSOCIATED WITH THE INAPPROPRIATE DEFIBRILLATION. THERE IS NO INFORMATION TO SUGGEST THE PATIENT SUSTAINED A SERIOUS INJURY. DEVICE EVALUATION WAS ACCOMPLISHED THROUGH A REVIEW OF THE PATIENT'S DOWNLOADED DATA FILE. REVIEW OF THE DATA DOES NOT INDICATE ANY DEVICE MALFUNCTION RELATED TO THE DEFIBRILLATION EVENT. MONITOR (B)(4) - REUSE. ELECTRODE BELT: (B)(4): 06/01/2009 - REUSE. INAPPROPRIATE DEFIBRILLATIONS ARE ON ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. (B)(4). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF.
A US DISTRIBUTOR CONTACTED ZOLL ON (B)(6) 2015 TO REPORT THAT A PATIENT EXPERIENCED AN INAPPROPRIATE DEFIBRILLATION EVENT WHILE AT WORK. THE PATIENT WAS TREATED AT 19:51:05 ON (B)(6) 2015. MOTION ARTIFACT CONTRIBUTED TO THE FALSE DETECTION. A REVIEW OF THE EVENT INDICATES THAT THE RESPONSE BUTTONS WERE PRESSED AFTER THE TREATMENT WAS DELIVERED. THE PATIENT DID NOT SEEK MEDICAL ATTENTION AND CONTINUED USE OF THE LIFEVEST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 328144 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Other |