FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 4792621 · Received May 15, 2015

Report

Report Number
3004123209-2015-00504
Event Type
Malfunction
Date Received
May 15, 2015
Date of Event
April 29, 2015
Report Date
June 12, 2015
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K041067
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ROUTINE TESTING CONFIRMED THAT THIS DEVICE WAS INSTALLED BY THE CUSTOMER IN JANUARY 2011 AND PERFORMED TO SPECIFICATION UP TO THE 12TH JANUARY 2014. THE DATA OBTAINED FROM THE DEVICE SHOWED EVIDENCE THAT THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY, RESULTING IN MANUAL POWER-UPS OF 10 MINUTES IN DURATION OCCURRING BETWEEN (B)(6) 2014. THE PAD-PAK BECAME DEPLETED DURING THIS TIME. A FURTHER PAD-PAK WAS INSTALLED AND MULTIPLE MANUAL POWER UPS OF 10 MINUTES DURATION WERE AGAIN RECORDED IN THE HISTORY LOG BETWEEN THE (B)(6) 2015. IT IS REASONABLE TO CONCLUDE THAT DEVICES RETURNED FOR THE REPORTED FAULT MAY BE ATTRIBUTED TO MEMBRANE FAILURE. THE TEN MINUTE TIME OUTS WOULD SUGGEST A FAILING MEMBRANE. THE FAULT WAS WITNESSED DURING THE INVESTIGATION. THE FAULT COULD NOT BE REPLICATED WITH A NEW MEMBRANE FITTED. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE. (B)(4).

Description of Event or Problem · 1

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE WAS MALFUNCTIONED, AS THE DEVICE SWITCHES ON AUTOMATICALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
317983 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES, LTD PAD

Patients

Seq Age Sex Outcome Treatment
1