FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4792503 · Received April 29, 2015

Report

Report Number
1052693-2015-00660
Event Type
Malfunction
Date Received
April 29, 2015
Date of Event
April 4, 2015
Report Date
December 21, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT NOT YET RETURNED FOR EVAL.

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER HAD AN INACCURATE REFERENCE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. EXPECTED BLOOD GLUCOSE RESULTS NON-FASTING RANGE FROM 80 TO 142MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. COMPARING BLOOD GLUCOSE TEST RESULTS TO SOLISV2 METER AND CLAIMS RESULTS ARE 40MG/DL HIGHER. BLOOD TEST PERFORMED USING BOTH METERS AND TEST RESULT WAS 120MG/DL WITH SOLISV2 METER AND 160MG/DL WITH TRUERESULT METER. BACK TO BACK BLOOD TEST PERFORMED DURING CALL NON-FASTING ((B)(6) 2015; 12:05PM) WITH RESULTS OF 119MG/DL 119MG/DL. VERIFIED STORAGE OF TEST STRIPS IS WITHIN INSTRUCTED SPECIFICATION. TEST STRIP LOT MFR'S EXPIRATION DATE IS 10/31/2017 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED NON-FASTING FROM METER MEMORY: 45MG/DL, (B)(6) 2015, 11:18AM; 160MG/DL, (B)(6) 2015, 11:04AM. ADVERSE EVENT NOT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF HIGH BLOOD RESULTS. EXPECTED BLOOD GLUCOSE RESULTS NON-FASTING RANGES FROM 80 TO 142MG/DL. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. COMPARING BLOOD GLUCOSE TEST RESULTS TO SOLISV2 METER AND CLAIMS RESULTS ARE 40MG/DL HIGHER. BLOOD TEST PERFORMED USING BOTH METERS AND TEST RESULT WAS 120MG/DL WITH SOLISV2 METER AND 160MG/DL WITH TRUERESULT METER. BACK TO BACK BLOOD TEST PERFORMED DURING CALL NON-FASTING ((B)(6) 2015; 12:05PM) WITH RESULTS OF 119MG/DL AND 119MG/DL. VERIFIED STORAGE OF TEST STRIPS IS WITHIN INSTRUCTED SPECIFICATION. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 10/31/2017 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED NON-FASTING FROM METER MEMORY: (B)(6). ADVERSE EVENT NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281099 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PR2158

Patients

Seq Age Sex Outcome Treatment
1 0 YR