FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4792459 · Received April 29, 2015

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
1052693-2015-00659
Event Type
Malfunction
Date Received
April 29, 2015
Date of Event
April 3, 2015
Report Date
December 21, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER HAS HIGH GLUCOSE VALUE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Additional Manufacturer Narrative · 1

PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).

Description of Event or Problem · 1

CONSUMER COMPLAINT OF "HI" BLOOD RESULTS. EXPECTED BLOOD GLUCOSE RESULTS NOT VERIFIED. CUSTOMER FELLS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. BACK TO BACK BLOOD TEST NOT PERFORMED DURING CALL. VERIFIED STORAGE OF TEST STRIPS IS WITHIN INSTRUCTED SPECIFICATION. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/15/2017 AND OPEN VIRAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED NON-FASTING FROM METER MEMORY: MEMORY 1: 120MG/DL, (B)(6) 2015, 05:04PM; MEMORY 2: "HI", (B)(6) 2015, 04:54PM, MEMORY 3: "HI" (B)(6) 2015, 04:53PM; MEMORY 4: 120MG/DL, (B)(6) 2015, 06:04PM;L MEMORY 5: 84MG/DL, (B)(6) 2015, 08:01PM. ADVERSE EVENT NOT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF "HI" BLOOD RESULTS. EXPECTED BLOOD GLUCOSE RESULTS NOT VERIFIED. CUSTOMER FEELS WELL AND OBSERVED NO SYMPTOMS. MEDICAL INTERVENTION IS NOT REQUIRED AT THIS TIME. BACK TO BACK BLOOD TEST NOT PERFORMED DURING CALL. VERIFIED STORAGE OF TEST STRIPS IS WITHIN INSTRUCTED SPECIFICATION. TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/15/2017 AND OPEN VIAL DATE IS (B)(6) 2015. RECALL TEST RESULTS PERFORMED NON-FASTING FROM METER MEMORY: (B)(6). ADVERSE EVENT NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281914 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PR2069

Patients

Seq Age Sex Outcome Treatment
1 0 YR