ANGIO-SEAL DEVICE
Report
- Report Number
- 3003681312-2015-00041
- Event Type
- Injury
- Date Received
- May 22, 2015
- Report Date
- April 29, 2015
- Manufacturer
- ST. JUDE MEDICAL PUERTO RICO, B.V.
- Product Code
- MGB
- PMA / PMN Number
- P930038
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE DEVICE HISTORY RECORD WAS NOT POSSIBLE SINCE THE BATCH NUMBER WAS UNAVAILABLE. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED. THE ANGIO-SEAL DEVICE INSTRUCTION FOR USE (IFU) CAUTIONS, SHOULD ISCHEMIC SYMPTOMS APPEAR, TREATMENT OPTIONS INCLUDE THROMBOLYSIS, PERCUTANEOUS EXTRACTION OF THE ANCHOR OR FRAGMENTS, OR SURGICAL INTERVENTION.
THE FOLLOWING WAS PUBLISHED IN THE ARTICLE, ¿LATE ONSET IATROGENIC LIMB ISCHAEMIA AFTER DEPLOYMENT OF AN ANGIO-SEAL VASCULAR CLOSURE DEVICE,¿ WHICH WAS PUBLISHED ONLINE TO THE BMJ CASE REPORTS 2015 ON 26 APRIL 2015: TWO PATIENTS UNDERWENT AND HAD SUBSEQUENT VASCULAR ACCESS SITE CLOSURE WITH ANGIO-SEAL VASCULAR CLOSURE DEVICES. THE DEVICES WERE DEPLOYED SUCCESSFULLY AT THE RIGHT COMMON FEMORAL ARTERY (CFA) ACCESS SITE WITH SUCCESSFUL HEMOSTASIS. THREE WEEKS LATER, BOTH PATIENTS PRESENTED WITH FEATURES OF LIMB ISCHAEMIA NEEDING FURTHER DIAGNOSTIC WORK-UP INCLUDING REPEAT ANGIOGRAPHY, WHICH REVEALED SUBTOTAL OCCLUSION OF RIGHT COMMON FEMORAL ARTERY AT THE LEVEL OF PRIOR ACCESS AND ANGIO-SEAL DEPLOYMENT SITE. BOTH THE PATIENTS UNDERWENT SUCCESSFUL SURGICAL REPAIR WITH RESTORATION OF DISTAL FLOW AND RESOLUTION OF SYMPTOMS. DOI: 10.1136/BCR-2015-209393. PLEASE REFERENCE MFR REPORT NUMBER 3003681312-2015-00043.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 336618 | ANGIO-SEAL DEVICE | DEVICE, HEMOSTASIS, VASCULAR | MGB | ST. JUDE MEDICAL PUERTO RICO, B.V. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |