FDA Adverse Event
Other
Summary report: N
ALM
MDR report key: 479217
·
Received August 18, 2003
Report
- Report Number
- 9710053-2003-00002
- Event Type
- Other
- Date Received
- August 18, 2003
- Date of Event
- July 10, 2003
- Report Date
- August 15, 2003
- Manufacturer
- ALM
- Product Code
- FSY
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
DURING A SURGICAL OPERATION, A PORTION OF A GLASS HEAT ABSORPTION FILTER BROKE AWAY FROM ITS SUBASSEMBLY, AND DROPPED OUT OF THE OPTICAL PATH. THE PIECE(S) WERE CONTAINED WITHIN THE LIGHTHEAD ASSEMBLY. THE SURGERY SESSION CONTINUED FOR APPROXIMATELY 3.5 HOURS WITH THIS BROKEN FILTER SUBASSEMBLY THAT ENCOMPASSES THE LIGHT SOURCE. THE HOSPTIAL STAFF NOTED THAT THE PATIENT'S SKIN AS PETECHIAE AND HAD A CIRCULAR BLISTERING TYPE BURN, ABOUT THE SIZE OF A SILVER DOLLAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALM | SURGICAL LIGHT | FSY | ALM | PRX4401 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |