Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED ON (B)(6) 2015 THAT THE SUBJECT ICD WAS EXPLANTED BECAUSE THE RRT POINT HAS BEEN REACHED.

Additional Manufacturer Narrative · 1

THE DEVICE MODEL INVOLVED IN THIS MDR REPORT IS NOT APPROVED IN THE UNITED STATES; HOWEVER, IT IS SIMILAR TO A DEVICE MANUFACTURED BY SORIN THAT WAS CLEARED OR APPROVED BY FDA FOR MARKETING IN THE UNITED STATES.

Description of Event or Problem · 1

THE PATIENT WAS ADMITTED IN HOSPITAL DUE TO SEVERAL INAPPROPRIATE SHOCKS. INTERROGATION OF THE DEVICE REVEALED A VENTRICULAR LEAD IMPEDANCE VALUE LOWER THAN 200 OHMS. THE PHYSICIAN DECIDED TO PERFORM A LEAD REVISION ON (B)(6) 2015 AND THE RV LEAD (ISOLINE) WAS ABANDONED. THE ICD REMAINS IMPLANTED.

Description of Event or Problem · 1

THE PATIENT WAS ADMITTED IN HOSPITAL DUE TO SEVERAL INAPPROPRIATE SHOCKS. INTERROGATION OF THE DEVICE REVEALED A VENTRICULAR LEAD IMPEDANCE VALUE LOWER THAN 200 OHMS. THE PHYSICIAN DECIDED TO PERFORM A LEAD REVISION ON (B)(6) 2015 AND THE RV LEAD (ISOLINE) WAS ABANDONED. THE ICD REMAINS IMPLANTED.

Description of Event or Problem · 1

THE PATIENT WAS ADMITTED IN HOSPITAL DUE TO SEVERAL INAPPROPRIATE SHOCKS. INTERROGATION OF THE DEVICE REVEALED A VENTRICULAR LEAD IMPEDANCE VALUE LOWER THAN 200 OHMS. THE PHYSICIAN DECIDED TO PERFORM A LEAD REVISION ON (B)(6) 2015 AND THE RV LEAD (ISOLINE) WAS ABANDONED. THE RRT HAS BEEN REACHED AND SO THE SUBJECT ICD HAS BEEN REPLACED.