PARADYM
Report
- Report Number
- 1000165971-2015-00326
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- May 11, 2015
- Report Date
- May 19, 2015
- Manufacturer
- SORIN GROUP ITALIA S.R.L. - CRM FACILITY
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
IT WAS REPORTED ON (B)(6) 2015 THAT THE SUBJECT ICD WAS EXPLANTED BECAUSE THE RRT POINT HAS BEEN REACHED.
THE DEVICE MODEL INVOLVED IN THIS MDR REPORT IS NOT APPROVED IN THE UNITED STATES; HOWEVER, IT IS SIMILAR TO A DEVICE MANUFACTURED BY SORIN THAT WAS CLEARED OR APPROVED BY FDA FOR MARKETING IN THE UNITED STATES.
THE PATIENT WAS ADMITTED IN HOSPITAL DUE TO SEVERAL INAPPROPRIATE SHOCKS. INTERROGATION OF THE DEVICE REVEALED A VENTRICULAR LEAD IMPEDANCE VALUE LOWER THAN 200 OHMS. THE PHYSICIAN DECIDED TO PERFORM A LEAD REVISION ON (B)(6) 2015 AND THE RV LEAD (ISOLINE) WAS ABANDONED. THE ICD REMAINS IMPLANTED.
THE PATIENT WAS ADMITTED IN HOSPITAL DUE TO SEVERAL INAPPROPRIATE SHOCKS. INTERROGATION OF THE DEVICE REVEALED A VENTRICULAR LEAD IMPEDANCE VALUE LOWER THAN 200 OHMS. THE PHYSICIAN DECIDED TO PERFORM A LEAD REVISION ON (B)(6) 2015 AND THE RV LEAD (ISOLINE) WAS ABANDONED. THE ICD REMAINS IMPLANTED.
THE PATIENT WAS ADMITTED IN HOSPITAL DUE TO SEVERAL INAPPROPRIATE SHOCKS. INTERROGATION OF THE DEVICE REVEALED A VENTRICULAR LEAD IMPEDANCE VALUE LOWER THAN 200 OHMS. THE PHYSICIAN DECIDED TO PERFORM A LEAD REVISION ON (B)(6) 2015 AND THE RV LEAD (ISOLINE) WAS ABANDONED. THE RRT HAS BEEN REACHED AND SO THE SUBJECT ICD HAS BEEN REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334225 | PARADYM | DEFIBRILLATOR, AUTO. IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRO. (CRT-D) | NIK | SORIN GROUP ITALIA S.R.L. - CRM FACILITY | PARADYM RF SONR CRT-D 9770 | 2596 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |