FDA Adverse Event Malfunction Summary report: N

UNIVERSAL BATTERY CHARGE II

MDR report key: 4791833 · Received May 22, 2015

Report

Report Number
1719045-2015-10278
Event Type
Malfunction
Date Received
May 22, 2015
Date of Event
October 6, 2014
Report Date
October 6, 2014
Manufacturer
SYNTHES MONUMENT
Product Code
HSB
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO VARIOUS WORN ELECTRICAL COMPONENTS FROM NORMAL WEAROUT. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS OBSERVED THAT THE UNIVERSAL BATTERY CHARGER DEVICE WOULD NOT TURN ON AND HAD NO POWER. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335778 UNIVERSAL BATTERY CHARGE II BATTERY, REPLACEMENT, RECHARGEABLE HSB SYNTHES MONUMENT 4804

Patients

Seq Age Sex Outcome Treatment
1