HOMECHOICE PRO
Report
- Report Number
- 1416980-2015-22513
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- December 19, 2014
- Report Date
- April 29, 2015
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED AND THE EVALUATION IS COMPLETE. DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THIS ISSUE WAS DETERMINED TO BE THE TOTAL TIDAL ULTRAFILTRATION (UF) REMOVAL WAS SET TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 23:16:30. DURING NIGHT DRAIN CYCLE 15, THE PATIENT'S ULTRAFILTRATION READING WAS 1107ML, INDICATING THE HOME PATIENT DRAINED 857ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 335953 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |