FDA Adverse Event Malfunction Summary report: N

CASING FOR 14.4V BATTERY

MDR report key: 4791634 · Received May 22, 2015

Report

Report Number
3009450871-2015-11624
Event Type
Malfunction
Date Received
May 22, 2015
Date of Event
August 1, 2014
Report Date
August 21, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
MOQ
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE HAD CORROSION ON THE CONTACTS. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMMERSION DURING CLEANING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CASING FOR BATTERY DEVICE WAS NOT WORKING. THE EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO DELAYS IN THE SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE REPORTER STATED THAT THE EVENT OCCURRED ON (B)(6) 2014; HOWEVER, THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
336368 CASING FOR 14.4V BATTERY BATTERY, REPLACEMENT, RECHARGEABLE MOQ DEPUY SYNTHES POWER TOOLS 005552

Patients

Seq Age Sex Outcome Treatment
1