CASING FOR 14.4V BATTERY
Report
- Report Number
- 3009450871-2015-11624
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- August 1, 2014
- Report Date
- August 21, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- MOQ
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER
Narratives
(B)(6). THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE HAD CORROSION ON THE CONTACTS. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMMERSION DURING CLEANING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT THE CASING FOR BATTERY DEVICE WAS NOT WORKING. THE EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO DELAYS IN THE SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE REPORTER STATED THAT THE EVENT OCCURRED ON (B)(6) 2014; HOWEVER, THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 336368 | CASING FOR 14.4V BATTERY | BATTERY, REPLACEMENT, RECHARGEABLE | MOQ | DEPUY SYNTHES POWER TOOLS | 005552 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |