G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-14902
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- April 26, 2015
- Report Date
- April 27, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION; HOWEVER, THE DATA LOG WAS PROVIDED BY THE PATIENT. IT WAS DOWNLOADED AND REVIEWED ON 05/12/2015 AND CONFIRMED THE REPORTED EVENT OF INACCURATE BG VALUES. THE TRANSMITTER WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE TRANSMITTER WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION.
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT CONTINUOUS GLUCOSE MONITOR (CGM) INACCURACY COMPARED TO BLOOD GLUCOSE (BG) METER ON (B)(6) 2015. SENSOR WAS INSERTED ON (B)(6) 2015. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.