FDA Adverse Event Malfunction Summary report: N

CRIT-LINE BLOOD CHAMBER

MDR report key: 4791398 · Received May 21, 2015

Report

Report Number
2937457-2015-00959
Event Type
Malfunction
Date Received
May 21, 2015
Date of Event
April 1, 2015
Report Date
May 8, 2015
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
KOC
PMA / PMN Number
K935958
Removal / Correction Number
Z-1716-2015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. A PRODUCT RECALL HAS BEEN INITIATED BY THE MANUFACTURER AND THE REPORTED PRODUCT ISSUE IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA.

Additional Manufacturer Narrative · 1

PLANT INVESTIGATION HAS NOT YET BEEN COMPLETED. A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

A HEMODIALYSIS USER FACILITY HAS REPORTED THAT DURING TREATMENT A BLOOD LEAK OCCURRED. THE LEAK WAS VISUALLY OBSERVED TO BE AN EXTERNAL LEAK LOCATED AT THE THREADED BOTTOM END OF THE CRIT LINE BLOOD CHAMBER THAT ATTACHES TO THE DIALYZER. THE PATIENT HAD NO ADVERSE EFFECTS AND NO MEDICAL INTERVENTION WAS REQUIRED. ESTIMATED BLOOD LOSS WAS 5-10CCS. SAMPLE HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331204 CRIT-LINE BLOOD CHAMBER KOC FRESENIUS MEDICAL CARE NORTH AMERICA 14121306

Patients

Seq Age Sex Outcome Treatment
1