SPEEDBAND SUPERVIEW SUPER 7?
Report
- Report Number
- 3005099803-2015-01432
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- May 6, 2015
- Report Date
- May 6, 2015
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- FHN
- PMA / PMN Number
- K020824
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
RECEIVED ONE SPEEDBAND DEVICE FOR ANALYSIS. VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND SOME RESIDUE PRESENT INDICATING USE/HANDLING. THERE WAS NO ISSUE NOTED WITH THE EXTENSION TUBE. PHYSICAL EXAMINATIONS OF THE LIGATOR HEAD FOUND ALL SEVEN BANDS PRESENT AND WERE MOVED OUT OF POSITION. THE LIGATOR TEETH WERE DAMAGED. THE SUTURE WAS INTACT AND ATTACHED TO THE TRIP WIRE LOOP. THE TRIP WIRE WAS NOT SECURED IN THE HANDLE ASSEMBLY SLOT, WITH THE PROXIMAL LOOP RETRACTED INTO THE HANDLE POST. THE TRIP WIRE CRIMP WAS CAUGHT INSIDE THE HANDLE POST. THERE WAS NO EVIDENCE PRESENT THAT THE TRIP WIRE HAD BEEN SECURED. FUNCTIONAL EVALUATION OF THE HANDLE WAS PERFORMED BY TURNING THE HANDLE KNOB AT 180 DEGREES; NO ISSUE WAS NOTED WITH THE HANDLE ASSEMBLY. BASED ON THE CONDITION OF THE RETURNED DEVICE, THE COMPLAINT THAT THE BANDS FAILED TO DEPLOY COULD NOT BE FUNCTIONALLY VERIFIED; HOWEVER, VISUAL EVALUATION INDICATE THAT THE DEVICE MOST LIKELY FAILED TO DEPLOY THE BANDS DURING THE PROCEDURE. IT DOES NOT APPEAR THAT THE TRIP WIRE WAS CINCHED IN THE HANDLE SLOT AS INSTRUCTED IN THE DFU, WHICH IMPACTED THE DEPLOYMENT ACTIVITY OF THE BANDS. THEREFORE THE MOST PROBABLE ROOT CAUSE CLASSIFICATION IS USER ERROR. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED AND CONFIRMED THAT THIS DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS.
(B)(4). THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPEEDBAND SUPERVIEW SUPER 7 DEVICE WAS USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2015. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BANDS WOULD NOT DEPLOY. THE PROCEDURE WAS COMPLETED WITH ANOTHER SPEEDBAND SUPERVIEW SUPER 7 DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPEEDBAND SUPERVIEW SUPER 7 DEVICE WAS USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2015. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BANDS WOULD NOT DEPLOY. THE PROCEDURE WAS COMPLETED WITH ANOTHER SPEEDBAND SUPERVIEW SUPER 7 DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 335163 | SPEEDBAND SUPERVIEW SUPER 7? | LIGATOR, HEMORRHOIDAL | FHN | BOSTON SCIENTIFIC - SPENCER | M00542253 | 17797305 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |