PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2015-16618
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- May 4, 2015
- Report Date
- May 4, 2015
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, A CRACKED BELT CLIP SLOT, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A STAINED END CAP STICKER.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
A CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THEIR INSULIN PUMP ALARMED BUTTON ERROR. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS 70 MG/DL AT THE TIME OF THE CALL. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD HAVE LED TO THE ISSUE. THE CUSTOMER WAS INFORMED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. THE CUSTOMER SENT THE PRODUCT BACK FOR ANALYSIS. A REPLACEMENT PUMP WAS SHIPPED TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 336740 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754LWWS |