PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-59142
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- May 4, 2015
- Report Date
- May 4, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
NO UNEXPECTED BUTTON ERROR ALARMS NOTED. THE INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE ON THE DOWN ARROW BUTTON DUE TO CORRODED KEYPAD TRACES NOTED. THE INSULIN PUMP HAD A CRACKED LCD WINDOW, CRACKED CASE AT LCD WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, SCRATCHES ON RESERVOIR TUBE WINDOW AND CRACKED BATTERY TUBE THREADS.
CUSTOMER REPORTED THE INSULIN PUMP FELL IN WATER AND IT ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE WAS 112 MG/DL. THE DEVICE ALARMED DURING NORMAL USE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE, REVERT TO BACK UP PLAN, AND DEVICE NEED TO BE REPLACED. CUSTOMER AGREED TO RETURN THE DEVICE FOR FURTHER ANALYSIS.