OT VERIO IQ METER
Report
- Report Number
- 2939301-2015-20709
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Report Date
- April 30, 2015
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE TEST STRIPS WERE EVALUATED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED; HOWEVER, A SECONDARY ISSUE WAS NOTED WHERE ERROR 4 WAS OBSERVED DURING CONTROL SOLUTION TESTS. THE CONTROL SOLUTION PASSED ALL TESTING WITH NO FAULTS FOUND. THE CONTROL SOLUTION WAS ALSO TESTED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN (B)(4) ALLEGING THAT THE SUBJECT METER READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE. THE REPORTER CLAIMED OBTAINING A BLOOD GLUCOSE READING OF "136MG/DL" WITH THE SUBJECT METER AND "71MG/DL" ON ANOTHER DEVICE, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 336014 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3779295 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |