RESERVOIR 1.8ML
Report
- Report Number
- 3004209178-2015-59067
- Event Type
- Malfunction
- Date Received
- May 22, 2015
- Date of Event
- May 4, 2015
- Report Date
- May 4, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K001828
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE CUSTOMER STATED THAT THE RESERVOIR WAS LEAKING WHEN BEING REMOVED. THE CUSTOMER EXPLAINED THAT INSULIN WAS DRIPPING OUT OF THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 121 MG/DL. THE CUSTOMER CONFIRMED THE LEAK WAS AT THE BACK OF THE RESERVOIR AND THAT THERE WAS LIQUID IN THE RESERVOIR COMPARTMENT. THE CUSTOMER STATED THAT THE LEAK WAS PAST THE 2ND O-RING. IN REGARDS TO THE INSULIN PUMP, THE CUSTOMER COULD HEAR THE MOTOR TURNING, BUT THE PISTON WAS NOT MOVING. THE CUSTOMER AGREED TO RETURN THE PRODUCT FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 335706 | RESERVOIR 1.8ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-326A | H7916246 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |